Technical
Specifications |
|
Display |
| |
64 x
128
pixels,
reflective
LCD,
blue
on
yellow |
|
|
Operational
Modes |
| |
Standalone
without
any
PC
software
or
with
the
VT
for
Windows
PC
software. |
|
|
Gas
Types |
| |
Air,
N2,
N2O,
CO2,
O2,
N2O
bal
O2,
N2
bal
O2 |
|
|
Battery
Power
Supply |
|
Maximum
Over-Voltage: |
15 V
dc |
|
Input
Voltage
Range: |
9 V
dc |
|
Power
Consumption: |
< 70
mA |
|
Battery
Life: |
> 7
hours |
|
|
External
Power
Supply |
|
Maximum
Over-Voltage: |
264
V ac |
|
Input
Voltage
Range: |
100
to
240
V ac |
|
Input
Frequency
Range: |
50/60
Hz |
|
Output
Voltage: |
12 V
to
15 V |
|
Output
Current: |
1.2
A |
|
Fuse
Rating: |
N/A |
|
|
Low-Pressure
Port |
|
Maximum
Applied
Pressure: |
5
psi |
|
Operating
Pressure
(Differential): |
- 20
cmH2O
to
120
cmH2O |
|
Operating
Pressure
(Common-Mode): |
N/A |
|
Span
Accuracy: |
+ 2
% of
reading
or
1.5
mmHg |
|
Frequency
Response: |
> 10
Hz |
|
Resolution: |
0.1
mmHg |
|
Sample
Rate: |
100
Hz |
|
Fittings: |
Flow
connector
with
2
tubes
"T"
connected
to a
single
Luer
fitting |
|
Note: |
No
fluid
may
be
applied
to
port |
|
|
High-Pressure
Port |
|
Maximum
Applied
Pressure: |
125
psi |
|
Operating
Pressure: |
- 2
psi
to
100
psi |
|
Span
Accuracy: |
+ 2
% of
reading
or +
0.2
psig |
|
Frequency
Response: |
> 10
Hz |
|
Resolution: |
0.1
psi |
|
Sample
Rate: |
100
Hz |
|
Fittings: |
Single
port,
Luer
lock,
stainless
steel |
|
Note: |
No
fluid
may
be
applied
to
port |
|
|
Airway
Pressure |
|
Maximum
Applied
Pressure: |
5
psi |
|
Operating
Pressure: |
- 20
cmH2O
to
120
cmH2O |
|
Span
Accuracy: |
+ 2
% of
reading
or +
0.5
cmH2O |
|
Frequency
Response: |
> 25
Hz |
|
Resolution: |
0.1
cmH2O |
|
Sample
Rate: |
100
Hz |
|
Fittings: |
Internally
connected
to
flow-sensor
pressure
lines |
|
|
High-Flow
Port |
|
Maximum
Flow
Rate
(absolute
value): |
200
lpm |
|
Operating
Flow
Range: |
+/-
200
lpm |
|
Accuracy: |
+/-
3 %
of
reading
or
+/-
2 %
of
range |
|
Floor
for
Absolute
Accuracy: |
25
lpm |
|
Resolution: |
0.01
lpm |
|
Frequency
Response: |
> 25
Hz
or t10-90
< 40
ms |
|
Sample
Rate: |
100
Hz |
|
Dynamic
Resistance: |
< 2
cmH2O
@ 60
lpm |
|
Low-Flow
Dropout: |
2.5
lpm |
|
Breath-Detect
Threshold: |
4
lpm |
|
Volume
Range: |
>
+/-
60 l |
|
Tidal
Volume
Accuracy: |
+/-
3 %
of
reading
or
+/-
20
ml,
whichever
is
greater |
|
Fittings: |
15
mm
OD/ID,
1:40
conical
male |
|
|
Low-Flow
Port |
|
Maximum
Flow
Rate
(absolute
value): |
35
lpm |
|
Operating
Flow
Range: |
+ 25
lpm |
|
Accuracy: |
+ 3
% of
reading
or +
1%
of
range |
|
Floor
for
Absolute
Accuracy: |
3
lpm |
|
Resolution: |
0.01
lpm
flow
> 1
lpm |
|
Frequency
Response: |
> 25
Hz
or t10-90
< 40
ms |
|
Sample
Rate: |
100
Hz |
|
Dynamic
Resistance: |
<
2.5
cmH2O
@ 5
lpm |
|
Low-Flow
Dropout: |
0.24
lpm |
|
Breath-Detect
Threshold: |
1
lpm |
|
Volume
Range: |
>+/-
60 l |
|
Volume
Accuracy: |
+ 3
% of
reading
or +
2 ml |
|
Fittings: |
15
mm
OD/ID,
1:40
conical
male |
|
|
Oxygen
Measurement |
|
Range: |
0 to
100
% |
|
Accuracy: |
+ 2
%
full-scale
output |
|
Resolution: |
0.1
% O2 |
|
Frequency
Response: |
> 15
s (t10-90) |
|
Sample
Rate: |
100
Hz |
|
Calibration: |
Galvanic
fuel
cell |
|
Notes: |
Automatic
partial
pressure
compensation
for
barometric
and
airway
pressure
changes |
| |
Recommended
interval
for
changing
oxygen
sensor
is
one
year.
However,
sensor
may
last
longer.
During
user
calibration
of
the
sensor,
the
VT
MOBILE
can
detect
if
the
sensor
needs
to
be
replaced. |
|
|
Barometric
Pressure
Measurement |
|
Range: |
8 to
18
psia
(400
mmHg
to
900
mmHg) |
|
Accuracy: |
+ 2
% of
reading |
|
Resolution: |
0.1
mmHg |
|
Frequency
Response: |
< 5
s (t10-90) |
|
Sample
Rate: |
N/A |
|
Calibration: |
Not
required;
however,
device
allows
user
calibration
of
offset |
|
|
Inspiratory
and
Expiratory
Tidal
Volume |
|
Resolution: |
0.1
ml |
|
Range: |
> 10
l |
|
Accuracy: |
± 3
%
expiratory
minute
volume |
|
Resolution: |
0.001
lpm |
|
Range: |
0 to
60 l |
|
Accuracy: |
± 3
% |
|
|
Breath
Rate |
|
Resolution: |
0.1
BPM |
|
Range: |
2
BPM
to
150
BPM |
|
Accuracy: |
± 1
%
inspiratory-to-expiratory
time
ratio
(I:E
ratio) |
|
Resolution: |
0.01
Range:
0.25
to
9.99 |
|
Accuracy: |
± 2
% or
0.1
s |
|
|
Peak
Inspiratory
Pressure |
|
Resolution: |
0.1
cmH2O |
|
Range: |
±
120
cmH2O |
|
Accuracy: |
+ 3
% or
1
cmH2O |
|
|
Inspiratory
Pause
Pressure |
|
Resolution: |
0.1
cmH2O |
|
Range: |
±
120
cmH2O |
|
Accuracy: |
+ 3
% or
1
cmH2O |
|
|
Mean
Airway
Pressure |
|
Resolution: |
0.1
cmH2O |
|
Range: |
+ 80
cmH2O |
|
Accuracy: |
+ 3
% or
0.5
cmH2O |
|
|
Positive-End
Expiratory
Pressure
(PEEP) |
|
Resolution: |
0.1
cmH2O |
|
Range: |
- 5
to
40
cmH2O |
|
Accuracy: |
+ 3
% or
0.5
cmH2O |
|
|
Peak
Expiratory
Flow |
|
Resolution: |
0.01
lpm |
|
Range: |
0
lpm
to
150
lpm |
|
Accuracy: |
± 3
% or
2 %
of
range |
|
|
Peak
Inspiratory
Flow |
|
Resolution: |
0.01
lpm |
|
Range: |
0
lpm
to
150
lpm |
|
Accuracy: |
± 3
% or
2 %
of
range |
|
|
Temperature |
|
Resolution: |
0.1
˚C |
|
Range: |
0 ˚C
to
50
˚C |
|
Accuracy: |
± 1
˚C |
|
Units: |
˚C,
˚F,
˚K |
|
|
Humidity |
|
Resolution: |
0.1% |
|
Range: |
0 %
to
100% |
|
Accuracy: |
± 5
% |
|
|
RS232
Serial
Communications |
| |
4-pin
modular
connector
located
on
upper-left
side
of
panel
RS-232
compatible
with
the
VT
Plus
for
Windows
software
application
(version
2.01.00
or
higher.) |
|
|
Environmental
Specifications |
|
Operating
Temperature: |
50
to
104
˚F
(10
to
40
˚C) |
|
Storage
Temperature: |
-13
to
122
˚F
(-25
to
50
˚C) |
|
Operating
Humidity: |
0 to
80 %
non-condensing
at
temperatures
to
31
˚C,
decreasing
linearly
to
50 %
relative
humidity
at
104
˚F
(40
˚C) |
|
Storage
Humidity: |
0 %
to
95 %
non-condensing |
|
Operating
Barometric: |
7
psia
to
18
psia |
|
Storage
Barometric: |
-1000
to
10000
ft
(787.9
mmHg
to
522.7
mmHg) |
|
|
Dimensions
(WxDxH) |
| |
10
cm x
20
cm x
3.8
cm
(4
in x
8 in
x
1.5
in) |
|
|
Weight |
|