   PS420 Multiparameter Patient Simulator


Overview | Features | Models | Standard Accessories | Optional Accessories | Specifications | Get A Quote
 




 

 
 
Overview

 


The PS420 Patient Simulator is a handheld, high-performance simulator for testing patient monitors.

Small enough to fit in a pocket, the handy PS420 features a wide variety of simulation capability, including a full range of ECG, respiration, blood pressure, temperature and cardiac output conditions. The tool includes 12-lead ECG, two-channel blood pressure simulation, 35 arrhythmia selections, pacemaker simulation as well as adult and pediatric normal sinus rhythms.

For convenient use, labelled hot keys on the keypad guide users to the most common settings.

 

 


 
Features


» Compact, lightweight, pocketsize
»
Labeled hot keys for common settings
»
12-lead ECG
»
Respiration and temperature selection
»
Two-channel blood pressure simulation
»
Optional cardiac output
»
Adult and pediatric normal sinus rhythms
»
35 arrhythmia selections
»
ECG performance waveformst
»
ST segment levels
»
ECG artifact
»
Pacemaker simulation
» RS232 serial port for computer control
» Battery operated

 

Models

 


» PS420
 
Standard Accessories

» Operator's manual
 
Optional Accessories


» 
 
Specifications


Specifications
ECG
Normal Rate: 80 BPM
Selectable Rates: 30, 40, 60, 80, 100, 120, 140, 160, 180,
  200, 220, 240, 260, 280, 300 BPM
Accuracy: + 1 %
Output Impedance: 500, 1000, 1500, and 2000 Ω for Leads I,
  II, and III
ECG Amplitudes: 0.5 mV, 1 mV, 1.5 mV, and 2 mV
Amplitude Accuracy: + 2 % Lead II
High-Level Output: 1000x Lead II
Waveform Type: Adult or Pediatric ECG Waveform

Performance Waveform
Lead II Square Wave:
  2 Hz, 0.125 Hz
Pulse:
  30, 60, 120 BPM, 60 ms pulse width
Sine Wave:
  0.5, 4, 10, 40 50, 60 Hz (1 mV amplitude, lead II)
Triangle Wave:
  2 Hz

ST Segment Analysis
Elevated or Depressed:
  0.8 mV to + 0.8 mV in 0.1 mV steps

Pacemaker
Pacer Spike Amplitude:
  2, 4, 8, and 10 mV in Lead II
Accuracy:
  + 5 %, lead II
Pacer Spike Duration:
  0.1, 0.5, 1, 1.5, and 2 ms
Accuracy:
  + 5 %
Selections:
  Asynchronous Pacemaker
  Pacer Non-Function
  Pacer Non-Capture
  Demand Occasional Sinus
  Demand Frequent Sinus
  AV Sequential

Blood Pressure
Input/Output Impedance:
  350 Ω
Exciter Input Limit:
  + 10 V
Exciter Input Frequency Range
  : DC to 4000 Hz
Transducer Sensitivity:
  5 or 40 µV/V/mmHg
Level Accuracy:
  + 1 %, + 1 mmHg
Static Levels BP1:
  - 10, 0, 50, 100, 150, 200, 250 mmHg
Static Levels BP2:
  - 10, 0, 80, 160, 240, 320, 400 mmHg
Channel Selections:
  Arterial 120/80, Channel 1 and 2
  Radial Artery 120/80, Channel 1 and 2
  Left Ventricle 120/00, Channel 1 and 2
  Right Ventricle 25/00, Channel 1 and 2
  Central Venous 15/10, Channel 2
  Pulmonary Artery 25/10, Channel 2
  Pulmonary Wedge 10/2, Channel 2
  Left Atrium 14/4; Automatic Swan/Ganz (every 20 sec)
  Manual Swan/Ganz (changes when entry is selected), Channel2
Note:
  Synchronized with all normal sinus rates. Physiologically track all arrhythmia selection

Cardiac Output
must have optional Cardiac Output Adapter Box p/n 2462200
Catheter Type: Baxter Edwards, 10 cc
Blood Temperature: 98.6 ˚F (37 ˚C)
CO for 35.6 ˚F (2 ˚C): 3, 5, 7 l/min
CO for 68 ˚F (20 ˚C): 3, 5, 7 l/min
Cal Pulse: Of 1 ˚C for 1 sec; of Delta 402 Ω for 4 sec.
Computational Constant: For 35.6 ˚F (2 ˚C) is 0.561; for 68 ˚F (20 ˚C) is 0.608
Left to Right Shunt: 35.6 and 68 ˚F (2 and 20 ˚C)
Faulty Injectate: 35.6 and 68 ˚F (2 and 20 ˚C)
Accuracy: + 5 %
Notes: Calibrated or uncalibrated cardiac output waves for 4 different CO values.

Respiration
Baseline Impedance:
  500, 1000, 1500, and 2000 Ω, Leads I, II,
  and III
Lead Selections:
  LL or LA
Impedance Variations
  3, 1, 0.5, and 0.2 Ω
Accuracy:
  + 5%
Rates:
  15, 20, 30, 40, 60, 80, 100, 120, and 0
  BPM for Apnea
Accuracy:
  + 2 %
Apnea:
  12 seconds, 22 seconds, 32 seconds, and
  continuous

Temperature
Compatibility:
  YSI 400/700 Series
Temperature:
  86, 95, 98.6, 104, and 107.6 ˚F (30, 35, 37,
  40 and 42 ˚C)
Temperature Simulation Accuracy:
  + 0.25 ˚C

Arrhythmias
 
  Base Rate of 80 BPM
  Sinus Arrhythmia
  Atrial (PAC)*
  Missed Beat*
  Atrial Tachycardia
  Atrial Flutter
  Nodal (PNC)*
  Nodal Rhythm
  Supraventricular Tachycardia
  PVC1 Left Ventricular Focus*
  PVC 1 Early, LV Focus *
  PVC1 R on T, LV Focus*
  PVC2 Right Ventricular Focus*
  PVC2 Early, RV Focus*
  PVC2 R on T, RV Focus*
  Multifocal PVCs*
  Atrial Fibrillation Coarse/Fine
  PVCs 6/minute
  PVCs 12/minute
  PVCs 24/minute
  Frequent Multifocal PVCs
  Bigeminy
  Trigeminy
  Pair PVCs*
  Run 5 PVCs *
  Run 11 PVCs*
  Ventricular Tachycardia
  Ventricular Fibrillation Coarse/Fine
  Asystole
  Conduction Defects
  First Degree
  Second Degree
  Third Degree
  Right Bundle Branch Block
  Right Bundle Branch Block
  *Will go to NSR ECG @ 80 BPM after completion

Artifacts
 
  50/60 Hz
  Muscle
  Baseline
  Respiration

General Information
Power:
  9 V battery/battery eliminator
Housing:
  ABS plastic case
Display:
  2-line by 16-character LCD with keypad
RS232:
  Bidirectional interface, 9600 Baud
Dimensions:
  6.1 in L x 3.7 in W x 1.3 in H ( 15.6 cm L
  x 9.4 cm W x 3.4 cm H)
Weight:
  0.9 lb (0.4 kg)
Temperature Requirements:
  Operating: 59 ˚F to 95 ˚F (15˚C to 35 ˚C)
  Storage: 32 ˚F to 122 ˚F (0 ˚C to 50 ˚C)

 