   PS415 Multiparameter Patient Simulator


Overview | Features | Models | Standard Accessories | Optional Accessories | Specifications | Get A Quote
 




 

 

 

 

 

 

 
 
Overview

 

The PS415 is a high-performance multiparameter patient simulator for testing full-function monitors. Compact and lightweight, the handheld PS415 provides the versatile features and convenience to meet the needs of mobile field service technicians.

The device features 12-lead ECG output, two invasive blood-pressure channels, a wide range of arrhythmia selections, and respiration and temperature simulation. With its affordable price and broad capability, the PS415 provides one of the best cost-benefit ratios among simulators on the market today.

Technicians control the PS415 with manual commands or onboard autosequences. A convenient pullout card provides quick menu access for all the manual simulations, and the front panel features an easy-to-use soft-key interface
 


 
Features


» Handheld
» 12-lead ECG simulations
» Wide range of arrhythmia selections
» Automatic or manual operation
» RS232 port for computer control
» Two invasive blood pressure channels
» Respiration simulation
» Temperature simulation
» Battery operated
» Pullout card for quick menu access
 
Models

 


» PS415
 
Standard Accessories

» Operator's manual
 
Optional Accessories


» 
 
Specifications


Specifications
Normal Sinus Rhythms
Rates: 30, 40, 60, 70, 80, 90, 100, 120, 140, 150, 160, 180, 200, 210, 220, 240, 270, 300, and 350 BPM
Accuracy: + 1 %
Auto-Step Rate: 30, 60, 80, 120, 180, and 240 BPM at 30-second intervals
Amplitude Lead II: 0.5, 1, 1.5, and 2 mV
Amplitude Accuracy: + 5 % 2-Hz square wave at 1 mV p-p (Lead II)
Limb-Lead Impedance Selections: 500 or 1000 Ω
V-Lead Impedance: 1000 Ω
Accuracy: + 5 %

ECG Performance
Manual Selections
Pulse: 4 seconds
Square Wave: 2 Hz
Triangle Wave: 2 Hz
Sine Waves: 0.5, 10, 40, 50, 60, and 100 Hz
Automated Sequence: Square wave; pulse; sine waves 10, 40, 50, 60, and 100 Hz; and triangle wave

ECG Arrhythmia Selections
Atrial fibrillation
   
Second-degree A-V block, type 1
   
Right-bundle-branch block
   
Premature atrial contraction
   
Premature ventricular contraction, standard
   
Premature ventricular contraction, early
   
Premature ventricular contraction, R on T
   
Multifocal PVCs
   
Bigeminy
   
Run of 5 PVCs
   
Ventricular tachycardia
   
Ventricular fibrillation
   
Ventricular fibrillation
   
Pacemaker
   
Fetal/Maternal
   

Blood Pressure Channels 1 and 2
Input/Output Impedance:
  300 Ω
Exciter Input Voltage Range:
  2 to 16 V
Exciter Input Frequency Range:
  DC to 4 kHz
Output Sensitivity Selections:
  5 or 40 µV/V/mmHg
Output Range:
  0 to + 300 mmHg
Accuracy:
  + 1 % of full range + 1 mmHg
Dynamic Blood Pressure Waveforms:
  Specified only at 80 BPM normal sinus rhythm ECG.
Rates:
  All dynamic pressures track all normal sinus rates and all arrhythmias.
Isolation:
  Electrically isolated from the rest of the instrument

BP Channel-1 Waveforms (mmHg)
Manual Selections:
Arterial: 120/80
Left Ventricle: 120/0
Central Venous Pressure: 15/10
Right Ventricle: 25/0
Pulmonary Artery: 25/10
Pulmonary Artery Wedge: 10/2
Atmosphere: 0
Static Pressures: 20, 40, 80, 100, 200, 250, and 300
Artifact: BP1/BP2
Automated Sequence: Atmosphere: 0, 20, 40, 80, 100, 200, 250, and 300

BP Channel-2 Waveforms (mmHg)
Atmosphere:
  0
Track BP1 Mode:
  BP 2 tracks BP1 setting to facilitate testing
  patient monitoring equipment with dual
  BP channels
Central Venous Pressure:
  15/10
Swan-Ganz:
  Insert
  Inflate
  Deflate
  Remove
Automated Sequence:
  Atmosphere: 0, 20, 40, 80, 100, 200, 250, and 300

Respiration
Output Configurations:
  Lead I, II, and RL to LL
Baseline Impedances:
  500 and 1000 Ω
Accuracy:
  + 5%
Delta Impedances:
  0.1, 0.2, 0.5, 1, and 3 Ω
Accuracy:
  + 10 % of setting
Normal Physiological Rates:
  0, 15, 20, 30, 40, 60, 80, 100, and 120 BPM
Accuracy:
  + 5 % of setting
Apneas:
  12 and 32 seconds

Temperature
Selections:
  86 ˚F (30 ˚C), 98.6 ˚F (37 ˚C), and 104 ˚F (40 ˚C)
Accuracy:
  + 0.7 ˚F (0.4 ˚C)
Probe Compatibility:
  YSI 400 series and YSI 700 series
Isolation:
  Electrically isolated from the rest of the instrument; connector shell is to common ground

General Information
Power:
  9 V battery for 50-hour life or line-
  operated via battery eliminator
Output:
  7.7 VAC, 100 mA, unregulated
Display:
  2-digit LCD
Lead Test:
  If the resistance is less than 1 kΩ, the
  display flashes at the rate of 4 Hz
Data Interface:
  RS232 compatible, bidirectional port
  5-pin DIN female connector
  Baud Rate: 2400
Connector:
  2.5 mm, center (+)
Dimensions:
  6.3 in L x 4 in W x 1.5 in H
  (16 cm L x 10.4 cm W x 3.8 cm H)
Weight:
  0.8 lb (0.3 kg)
Temperature:
  Operating: 59 ˚F to 95 ˚F (15 ˚C to 35 ˚C
  Storage: 32 ˚F to 131 ˚F (0 ˚C to 55 ˚C)