   PS410 ECG/Arrythmia Simulator


Overview | Features | Models | Standard Accessories | Optional Accessories | Specifications | Get A Quote
 




 

 
 
Overview

 

The PS410 ECG/Arrythmia Simulator is a compact, high-performance simulator for patient monitor testing.

This handheld device simulates a full range of cardiac rhythms and a wide variety of ECG conditions. It includes pacemaker simulation, 35 arrhythmia selections, and adult and pediatric normal-sinus rhythms.

Small enough to fit in a pocket, the handy PS410 weighs less than a pound and is easy to operate. Technicians simply connect the simulator to the device under test and use the PS410 keypad to enter the code presets. The simulator then transmits the selected preset simulations to the device
.
 

 
Features


» Compact, lightweight, pocket-sized
» 
Easy top-panel control
» 
12-lead ECG
» 
Adult and pediatric normal-sinus rhythms
» 
35 arrhythmia selections
» ECG performance waveforms
» 
ST segment levels
» 
Pacemaker simulation
» ECG artifact
» 
Battery operated
 
Models

 


» PS410
 
Standard Accessories

» Operator's manual
 
Optional Accessories


» 
 
Specifications


Specifications
Normal Sinus Rhythm
  12-lead with independent outputs referenced to RL
Normal Rate: 80 BPM
Selectable Rates: 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM
Accuracy: + 1 %
Output Impedance: 940 Ω between leads
ECG Amplitudes: 0.5 mV, 1 mV, and 2 mV
Amplitude Accuracy: + 2 % Lead II
High-Level Output: 1000x Lead II
Waveform Types: Adult or Pediatric ECG Waveform

ECG Performance
Lead II
Square Wave: 2 Hz, 0.125 Hz
Pulse: 30, 60, 120 BPM; 60 ms pulse width
Sine Waves: 0.5, 5, 10, 40, 50, 60 Hz (1 mV amplitude only, lead II)
Triangle Wave: 2 Hz

ST Segment Analysis
Elevated or depressed:
  0.6 mV to 6 mV in 0.2 mV steps

Selections
Pacemaker
  Pacemaker rhythm
  Pacer non-capture*
  Pacer non-function
  Demand pacer with occasional sinus
  Demand pacer with frequent sinus
  A-V sequential
Artifact
  50 Hz artifact
  60 Hz artifact
  Muscle artifact
  Baseline artifact
  Respiration artifact
Arrhythmia
  PVC1 Left Ventricular Focus*
  PVC1 Early, LV Focus*
  PVC1 R on T, LV Focus*
  Pair of PVCs*
  Run of 5 PVCs*
  Run of 11 PVCs*
  Multi-focal PVCs*
  Frequent multi-focal PVCs*
  PVC2 Right Ventricular Focus*
  PVC2 Early, RV Focus*
  PVC2 R on T, RV Focus*
  Bigeminy
  Trigeminy
  PVCs at a rate of 6 per minute
  PVCs at a rate of 12 per minute
  PVCs at a rate of 12 per minute
  Ventricular Tachycardia
  Ventricular Fibrillation
  Ventricular Fibrillation 1/2 size
  Supraventricular Tachycardia
  Premature atrial contraction*
  Premature nodal contraction*
  Asystole
  Missed beat*
  Nodal Rhythm
  Irregular Rhythm
  Atrial flutter
  Atrial fibrillation
  Atrial fibrillation 1/2 size
  Atrial Tachycardia
  First-Degree Block
  Second-degree block
  Third-degree block
  Right-bundle-branch block
  Left-bundle-branch block
  * Event occurs once. To repeat, enter the selection again

Operation
 
  Outputs are easily selected or adjusted using top-panel controls and the large two-digit numeric display.

Controls
Display:
  2-line LCD
Control:
  6 control keys; ON/OFF power switch

General Information
Power:
  9 V battery/battery eliminator
Housing:
  ABS plastic case
Dimensions
  6.1 in L x 3.7 in W x 1.3 in H (11.3 cm L x 7 cm W x 3.4 cm H)
Weight:
  0.9 lb (0.4 kg)
Temperature:
  Operating: 59 ˚F to 95 ˚F (15˚C to 35 ˚C)
  Storage: 32 ˚F to 122 ˚F (0 ˚C to 50 ˚C)

 