    MedSim 300B

Overview | Features | Models | Standard Accessories | Optional Accessories | Specifications | Get A Quote
 

 
Overview

 

Comprehensive & Cost-Effective.

No other patient simulator on the market today offers you so much versatility. Designed to deliver simulations for your most complex inspection requirements, the medSim 300B offers a 12-lead ECG, four blood pressure channels, impedance-based respiration, temperature, and cardiac output (optional). As part of our Integrated System approach, medSim can be controlled remotely from a medTester 5000 Series Automated Safety and Performance Analyzer. You won’t find an easier way increase your test activities.

The medSim 300B has factory preset waveform sequences along with 15 programmable sequences that simulate a wide range of physiological events. medSim’s compact size and handy optional controller, make the industry’s most comprehensive patient simulator an ideal investment for any biomedical department.

12-Lead ECG
User Programmable normal sinus & PVC parameters
Full range of cardiac arrhythmias and performance testing selections
User-programmable pacer spike width & amplitudes
R-wave detector testing
Power line, muscle, & baseline artifacts

The medSim 300B’s independent signal channels replicate the heart’s electrical activity. These 12-Lead format signals test the full range of cardiac equipment, from simple portable monitors to complex diagnostic recorders. Thirty basic arrhythmias, along with practically any combination of premature beat and artifact, are available. All of the performance waveforms for gain, frequency response and linearity testing that you would expect are standard… plus unique R-Wave detection testing & pacer simulation capabilities.

Blood Pressure
4 independent channels with both dynamic simulation and static levels
Complete Swan-Ganz procedure
Two triangle waveforms for linearity testing
Selectable BP/respiratory artifact levels
Automated static level sequencing

medSim is the only simulator on the market to offer four independent BP channels and automated static pressure level sequencing for hands-free testing.

Respiration
Adult, neonatal, and ventilator selection
Selectable I/E ratios
Baseline shift
Apnea episodes

You can use the medSim 300B to simulate the widest range of respiratory signals. This gives you the capacity to test impedance-based cardiorespirograms (CRGs), which are incorporated into the most sophisticated neonatal monitoring systems.

Temperature
Dual channels
Trending sequences

Verify the operation of a monitor’s delta computation with medSim’s exclusive dual temperature channels. Other features include dynamic temperature with three physiological trends that ensure the functionality of both long- and short-term trending capabilities.

Cardiac Output Option
Fixed waveforms at 3, 5, and 7 liters per minute
Three trending sequences
Normal & defective curves

Pushbutton Ease of Use
Use the optional remote controller, for convenient access to all medSim 300B waveform selections. Simply plug the controller into the RS-232 port on medSim’s top panel.

Training Capabilities
One of medSim’s "bonus" features is the ability to use it as a training device for basic defibrillator operation and waveform recognition.

Defibrillator Training
With two emergency scenarios and a cardioversion procedure, the medSim 300B can be used for basic defibrillator training.

Adding Capability with Option 1
Take advantage of extended capabilities with Option 1 software, which gives you the following powerful inspection features:

Cardiac Catheterization
When equipped with Option 1, medSim 300B will simulate a pressure gradient range of 5% to 50% across the aortic, mitral, and pulmonary heart valves.

Intra-Aortic Balloon Pump
In this mode, medSim 300B will generate an augmented arterial waveform whenever an assist TTL level pulse is sensed at the sync input.

Fetal Maternal ECG & Interactive IUP
Using Option 1 in the fetal/maternal mode, you can output a composite ECG complex.

Arrhythmia Selections
The medSim 300B offers 32 arrhythmia selections plus a programmable PVC function for a broad range of test and training capabilities. Select any one of these arrhythmias using the optional controller or directly from the menu selections:

Supraventricular
Atrial Fibrillation 1 & 2
Atrial Flutter
Sinus Arrhythmia
Missed Beat 80 & 120 BPM
Paroxysmal Atrial Tachycardia
Nodal
Supraventricular Tachycardia

Premature Beats
Atrial (PAC)
Nodal (PNC)
PVC-1
Multifocal PVCs

 

Features


» 12-Lead ECG simulation
» 4 BP channels
» Respiration and temperature
» Optional cardiac output simulation
» R-wave detector testing
» Optional hand-held controller
» Option 1 expansion
» Training capabilities
» Interfaces with medTester 5000C
 

 
Models


 


» medSim 300B
 

 
Standard Accessories


» Battery Eliminator (CSA Approved)
» Soft Vinyl Carrying Case
» Operating & Service Manual
» Two 9-volt batteries

 
Optional Accessories


» medSim 300B Remote Controller
» BP Cable, unterminated
» BP Cable, prewired for select patient monitors
» Temperature Cable for YSI 400 Series
» Temperature Cable for YSI 700 Series

» RS-232 Cables (DB25):
    - Female, pin 2 is Tx
    - Male, pin 2 is Rx
    - Male, pin 2 is Tx

» Calibration Cables:
    - BP Calibration Cable
    - Temperature Calibration Cable
    - Cardiac Output Calibration Cable
 
Specifications



Sync
» Accepts input during defib training and IABP augmentation (Option 1)

RS-232 Interface
» With the exception of those functions controlled by slide switches, most functions may be controlled remotely. Non-programmable functions include all utilities. Optional RS-232 cable required for remote control via PC.

ECG General
» Lead Configuration: 12-Lead ECG with 9 independent outputs for each signal lead referenced to RL.
» Output Impedances: 500, 1000, 1500, & 2000 ohms to RL.
» High Level Output: 0.2 V/mV of low level selection.
» Amplitude Accuracy: ± 5% 2 Hz Square Wave @ 1.0 mV p-p (Lead II).

PVC Parameter Selection
» Premature Beat Insertion: Push-button insertion of PVC, PAC, and PNC during 30 to 160 BPM normal sinus rhythm.
» Types: 1,2,3,4 (Ventricular Sites)
» Timing: R on T, Early, & Standard. Per minute: 0-25 with manual and auto stepping modes. Auto step interval is 60 seconds.
» Count: Number of PVC’s generated, 65535 max.

Power
» Two 9-volt batteries; battery eliminator is included with medSim 300B purchase. U.S./Domestic orders are shipped with a standard U.S. 115-volt style eliminator. International shipments come with either a 115-volt or 230-volt eliminator, as specified at the time of ordering. International 230-volt eliminators are accompanied by an international configuration detachable power cord. These power cords are available for most countries.

Pacemaker
» Rhythms: Async 75 BPM, Demand 1, Demand 2, AV Seq, Non capture, Non function. Atrial and ventricular pacer pulses can be independently set.
» Pulse: -700 mV to +700 mV (29 levels). Accuracy : 10% of setting ± 0.2 mV
» Width: 0.1, 0.2, 0.5, 1.0, 2.0 ms. Accuracy is 5%. Both amplitude and width have manual and auto stepping modes. Auto step time interval is 6 seconds.
» Pulse Polarity: Positive or negative.

Arrhythmia Selections
» Baseline Rate: 80 BPM
» Supraventricular: Afib1, Afib2, Aflutter, Sinus Arr, Missed Beat 80, Missed Beat 120, PAT, Nodal, SVT
» Premature Beats: Atrial (PAC)
» Nodal (PNC), PVC-1, Multi-focalPVC’s
» Ventricular Rhythms: Frequent Multi-focal 24/minute, Bigeminy, Trigeminy, Pair PVC, Run 5 PVC’s, Run 11 PVC’s, Ventricular Tachycardia, V-FIB fine, V-FIB coarse, Dying Heart Complex (EMD), Asystole
» Conduction: First Degree, Second Degree, Types 1 & 2, Third Degree, RBBB, LBBB
» Pacemaker: See "Pacemaker" section

Normal Sinus Rhythm
» Rates: 30, 60, 80, 120, 160, 200, 240, 300 BPM. Accuracy ± 1%.
» Amplitudes (Lead II): 50 m V to 0.5 mV in 0.05 mV steps; 0.5 mV to 5.5 mV in 0.25 mV steps. Manual and automatic step interval is 6 secs.
» ST Segments: 18 total - normal (iso-electric), elevated & depressed.
» ST Segment Levels (Lead II): -0.8 mV to +0.8 mV in 0.1 mV steps on Lead II, including two ± 0.05 mV steps. Includes manual and auto stepping modes at 12 second intervals. Operates at ECG rates from 30 BPM to 160 BPM. Modifies baseline ECG during arrhythmias.
» Axis Deviation: Normal (intermediate), horizontal, and vertical. Modifies baseline ECG during arrhythmias.
» Neonatal Mode: ECG R wave width is reduced to 40ms and the dynamic BP selections are reduced by 25%.

ECG Performance Testing
» Amplitudes set by baseline ECG amplitude. Lead I is 0.7 x Lead II; Lead III is 0.3 x Lead II; V Leads are same as Lead II.
» Waveforms (Lead II and V Leads)
» Square Wave: 2 Hz @ 1.0 mV p-p
» Pulse: 4.0 secs @ +1.0 mV
» Sine Waves: 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz @ 1 mV.
» Triangle Wave: 2 Hz @ 1 mV
» R Wave Detector Test: 60 BPM haver-triangle wave with selectable amplitude and width that can be automatically or manually operated.
» Width: 8.0 ms to 200 ms (12 selections)
» Amplitude (Lead II and V Leads): 50 m V to 0.5 mV in 0.05 mV steps; 0.5 mV to 5.5 mV in 0.25 mV steps.
» Lead I: 0.7 of Lead II
» Lead III: 0.3 of Lead II

Dimensions
» 10" L x 7" W x 3" H
» (25.4 cm L x 17.8 cm W x 7.6 cm H)

Weight
» 3.5 lb (1.6 kg)
 

Back to top