Introducing the new MPS450.
The new MPS450 is Fluke Biomedical's next generation, portable, multiparameter patient simulator for your comprehensive testing and training needs. Whether it's a quick check on a bedside monitor, arrhythmia recognition training, or performing a complete PM on the latest patient monitoring systems, the most recent simulator from Fluke Biomedical is a clear choice with its broad range of physiological waveforms, easy-to-use interface and compact, portable size.
 

Features


» ECG Simulation / 12-lead independent outputs
» 43 arrhythmia selections
» Four Invasive BP channels including Swan-Ganz
» Respiration simulation
» Temperature simulation
» Cardiac output simulation
» BP synchronized with ECG
» Fetal / maternal ECG simulation with IUP activity
» Physiological patient data
» Large, bright 4-line by 20 character supertwist display
» R-Wave detection test
» 9 V battery operated
» AC battery eliminator
» External battery tests (9 V)
» RS232 port for computer control
» Compact size
» Front panel soft key user interface
» Universal ECG connectors
» High-level ECG output
» CE Mark
» Expansion port
 
Models

 


» MPS450:Multiparameter Patient Simulator
Includes: ECG Simulation 12-lead, invasive BP, respiration, temperature, BP in sync with ECG, Large, bright 4-line by 20-character display, Physiological patient data, R-wave detection test, RS232 port for computer control, improved user interface with soft key navigation method, universal ECG connectors, flash memory for easy program upgrade.

» MPS450 W/CO:Multiparameter Patient Simulator With Cardiac Output
Includes: Everything listed above for MPS450, plus cardiac output simulation and the associated cardiac output adapter box.

» MPS450 W/FE:Multiple Patient Simulator With Fetal Simulations
Includes: Everything listed above for MPS450, plus fetal simulations. Selectable fetal heart rates, maternal heart rate, dynamic intrauterine pressure waveform.

» MPS450 W/CO/FE:Multiparameter Patient Simulator with cardiac output simulations and fetal simulations.
Includes: Everything listed above for MPS450, MPS450 with Cardiac Output, and MPS450 with Fetal Simulations.
 

 

Standard Accessories


» Operators manual
» Cardiac output adapter box (for units with Cardiac Output)
 
Optional Accessories


» Battery eliminator(US or INT)
» Carrying case
» RS233 cable
» Service manual
» Temp. cables-400/700 Series YSI
» BP Cables (Also for use with BPPUMP & Lionheart 2, 3 & 3c)
   Burdick(10 socket)
   Care(5 pin)
   Corometrics(3 pin/3 socket)
   Corometrics(12 pin)
   Criticare(6 pin)
   Critikon(6 pin)
   Datascope(6 socket)
   Datex(10 socket)
   Gould/Statham(5 pin)
   Hewlett Packard(5 socket)
   Hewlett Packard/Merlin(12 pin)
   Invivo Research(6 pin)
   Ivy Biomed(6 pin)
   Kontron/Roche(6 socket)
   Marquette 7000(8 pin)
   Marquette Eagle(11 pin)
   Marquette Twin
   Medical Data Electronics(MDE)(6 pin)
   Mennen Medical(6 pin)
   MPS-1 Cable to Lionheart 3 Adapter(6 pin)
   Nihon Khoden(5 pin)
   North American Drager(6 pin)
   ovametrix(3 pin/3 socket)
   Ohmeda(6 pin)
   Physio-Control(6 pin)
   Protocol(6 pin)
   Puritan Bennett(10 socket)
   Siemens(10 pin)Use with the Siemens Medical
   Transducer Adapter(3368-383-E530U) to run a
   single invasive BP Channel on the Siemens Medical
   sSC6000 & SC9000 Series Monitors.

   SMEC(6 socket)
   SpaceLabs(6 pin)Use with the SpaceLabs adapters 700-
   0028-00 & 0120-0551-00 when testing UltraView
   Comm& Module

   SpaceLabs/Squibb(5 pin)
   Tektronix/Squibb(6 pin)
   Universal BP Adapter(Pigtail/Unterminated)
   Vitastat(5 pin)
   Vitatek/Squibb(6 pin)
Specifications


Pacemaker
Atrial
Asynchronous
Demand with frequent sinus beat
Demand with occasional sinus beat
A-V sequential
Non-capture non-function

Accuracy
All amplitudes +/- 2%
All rates +/- 1%
All widths +/- 1%

ST Segments
Elevated or depressed: -0.8mV to +0.8mV in 1mV steps

Cardiac Output (optional)
Faulty Injectate Curve
Left to Right Shunt Curve
C.O. For 0o: 2.5, 5, and 10 l/min
C.O. For 24o: 2.5, 5, and 10 l/min
Cal Pulse: 1o for 1 second

Respiration
Baseline Impedance: 500 to 2000 ohms, LEADS I, II, III
Impedance Variations: 3, 1, 0.5, 0.2 Ω
Rates: 15 to 120 and 0 rpm for APNEA
Apnea Selections: 12, 22, 32 seconds, and continuous

ECG Performance
Amplitude: 0.5 to 2.0 mV
Square Wave: 2.0, 0.125 Hz
Pulse: 30, 60 BPM, 60 mS pulse width
Sine Waves: 0.5, 4.0, 10, 40, 50, 60, and 100 Hz
Triangle Wave: 2.0, 2.5 Hz

Normal Sinus Rhythm
12-lead configuration with independent outputs
Amplitude: 0.1, 0.25, 0.5, 1.0, 2.0 mV
Rates: 30 to 300 BPM
ECG Waveform Selections: adult or pediatric
Superimposed Artifact: 50 and 60 Hz, muscle, baseline wander, respiration

Fetal / Maternal / IUP Simulations (optional)
Fetal heart rates:60, 90, 120, 140, 150, 210, & 240 BPM
Uniform deceleration:140-100-140 BPM. Rate change lags iup contraction by 30 sec
Uniform acceleration:140-175-140 BPM. Rate change lags iup contraction by 30 sec
Early deceleration:140-100-140 BPM. No iup lag time
Late deceleration:140-100-140 BPM. Starts at iup peak
Maternal heart rate fixed at 80 BPM
Waveform:12 lead ECG with complete p-qrs-t complex
Dynamic intrauterine pressure (iup) waveform:positive bell shaped pressure curve
Peak pressure:90 mmhg
Contraction duration:90 sec
Pressure transducer sensitivity:5 or 40 m v/v/mmhg
Input/output impedance:300 ?
Exciter input voltage range:2.0 VDC to 16.0 VDC
Exciter input frequency range: dc to 5 kHz
Contraction/rate selections:manual, 1contraction/2 min, 1 contraction/3 min, 1 contraction/5 min
Level accuracy:90 mmhg +/-4mmhg (max)

Arrhythmia Selections
Premature BeatsPremature Atrial Contraction (PAC)
Nodal Premature Nodal Contraction (PNC)
Premature Ventricular Contraction (PVC)1
Left Ventricular Focus
PVC1 Early, Left Ventrical (LV) Focus
PVC1 R-on-T, Left Ventrical (LV) Focus
PVC2 Right Ventricular Focus
PVC2 Early, RV Focus
PVC2 R-on-T RV Focus
Multifocal PVCs

Conduction Defects
First Degree Heart Block
Second Degree Heart Block
Third Degree Heart Block
Right Bundle Branch Block
Left Bundle Branch Block

Supraventricular Beats
Atrial Fibrillation (Coarse)
Atrial Fibrillation (Fine)
Atrial Flutter
Sinus Arrhythmia
Missed Beat (1 time event)
Paroxysmal Atrial Tachycardia
Nodal
Supraventricular Tachycardia
Ventricular Rhythm
PVCs 6/Minute
PVCs 12/Minute
PVCs 24/Minute
Frequent Multifocal
Bigeminy
Trigeminy
Pair PVCs (1 time event)
Run 5 PVCs (1 time event)
Run 11 PVCs (1 time event)
Ventricular Tachycardia
Ventricular Fibrillation (Coarse)
Ventricular Fibrillation (Fine)
Asystole

Blood Pressure Channels
Dynamic BP waveforms are synchronized with normal sinus rhythm rates and track arrhythmia selections.
Respiration artifact can be selected on blood pressure channels
Transducer Sensitivity: 5 or 40 µV/V/mmHg
Calibrated Rate: 80 BPM normal sinus rhythm
Static Levels BP1: -10, 80, 160, 240, 320, 400 mmHg
Static Levels BP2: -10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels BP3: -50, 20, 40, 60, 80, 100 mmHg
Static Levels BP4: track BP3 levels

P1 Channel Selections
Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right ventricle: 25/00

P2 Channel Selections
Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (Central Venous): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary wedge: 10/2
Left atrium: 14/4

P3 Channel Selections
Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary wedge: 10/2
Left atrium: 14/4

P4 Channel Selections
Automatic Swan-Ganz (every 15 seconds)
Manual Swan-Ganz, changes each time Enter is selected
Temperature
Temperatures: 0oC, 24oC, 37oC, and 40oC
 


 

       Back to top