Pacemaker
Atrial
Asynchronous
Demand with frequent sinus beat
Demand with occasional sinus beat
A-V sequential
Non-capture non-function
Accuracy
All amplitudes +/- 2%
All rates +/- 1%
All widths +/- 1%
ST Segments
Elevated or depressed: -0.8mV to +0.8mV in 1mV steps
Cardiac Output (optional)
Faulty Injectate Curve
Left to Right Shunt Curve
C.O. For 0o: 2.5, 5, and 10 l/min
C.O. For 24o: 2.5, 5, and 10 l/min
Cal Pulse: 1o for 1 second
Respiration
Baseline Impedance: 500 to 2000 ohms, LEADS I, II, III
Impedance Variations: 3, 1, 0.5, 0.2 Ω
Rates: 15 to 120 and 0 rpm for APNEA
Apnea Selections: 12, 22, 32 seconds, and continuous
ECG Performance
Amplitude: 0.5 to 2.0 mV
Square Wave: 2.0, 0.125 Hz
Pulse: 30, 60 BPM, 60 mS pulse width
Sine Waves: 0.5, 4.0, 10, 40, 50, 60, and 100 Hz
Triangle Wave: 2.0, 2.5 Hz
Normal Sinus Rhythm
12-lead configuration with independent outputs
Amplitude: 0.1, 0.25, 0.5, 1.0, 2.0 mV
Rates: 30 to 300 BPM
ECG Waveform Selections: adult or pediatric
Superimposed Artifact: 50 and 60 Hz, muscle, baseline wander, respiration
Fetal / Maternal / IUP Simulations (optional)
Fetal heart rates:60, 90, 120, 140, 150, 210, & 240 BPM
Uniform deceleration:140-100-140 BPM. Rate change lags iup contraction by 30 sec
Uniform acceleration:140-175-140 BPM. Rate change lags iup contraction by 30 sec
Early deceleration:140-100-140 BPM. No iup lag time
Late deceleration:140-100-140 BPM. Starts at iup peak
Maternal heart rate fixed at 80 BPM
Waveform:12 lead ECG with complete p-qrs-t complex
Dynamic intrauterine pressure (iup) waveform:positive bell shaped pressure curve
Peak pressure:90 mmhg
Contraction duration:90 sec
Pressure transducer sensitivity:5 or 40 m v/v/mmhg
Input/output impedance:300 ?
Exciter input voltage range:2.0 VDC to 16.0 VDC
Exciter input frequency range: dc to 5 kHz
Contraction/rate selections:manual, 1contraction/2 min, 1 contraction/3 min, 1 contraction/5 min
Level accuracy:90 mmhg +/-4mmhg (max)
Arrhythmia Selections
Premature BeatsPremature Atrial Contraction (PAC)
Nodal Premature Nodal Contraction (PNC)
Premature Ventricular Contraction (PVC)1
Left Ventricular Focus
PVC1 Early, Left Ventrical (LV) Focus
PVC1 R-on-T, Left Ventrical (LV) Focus
PVC2 Right Ventricular Focus
PVC2 Early, RV Focus
PVC2 R-on-T RV Focus
Multifocal PVCs
Conduction Defects
First Degree Heart Block
Second Degree Heart Block
Third Degree Heart Block
Right Bundle Branch Block
Left Bundle Branch Block
Supraventricular Beats
Atrial Fibrillation (Coarse)
Atrial Fibrillation (Fine)
Atrial Flutter
Sinus Arrhythmia
Missed Beat (1 time event)
Paroxysmal Atrial Tachycardia
Nodal
Supraventricular Tachycardia
Ventricular Rhythm
PVCs 6/Minute
PVCs 12/Minute
PVCs 24/Minute
Frequent Multifocal
Bigeminy
Trigeminy
Pair PVCs (1 time event)
Run 5 PVCs (1 time event)
Run 11 PVCs (1 time event)
Ventricular Tachycardia
Ventricular Fibrillation (Coarse)
Ventricular Fibrillation (Fine)
Asystole
Blood Pressure Channels
Dynamic BP waveforms are synchronized with normal sinus rhythm rates and track arrhythmia selections.
Respiration artifact can be selected on blood pressure channels
Transducer Sensitivity: 5 or 40 µV/V/mmHg
Calibrated Rate: 80 BPM normal sinus rhythm
Static Levels BP1: -10, 80, 160, 240, 320, 400 mmHg
Static Levels BP2: -10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels BP3: -50, 20, 40, 60, 80, 100 mmHg
Static Levels BP4: track BP3 levels
P1 Channel Selections
Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right ventricle: 25/00
P2 Channel Selections
Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (Central Venous): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary wedge: 10/2
Left atrium: 14/4
P3 Channel Selections
Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary wedge: 10/2
Left atrium: 14/4
P4 Channel Selections
Automatic Swan-Ganz (every 15 seconds)
Manual Swan-Ganz, changes each time Enter is selected
Temperature
Temperatures: 0oC, 24oC, 37oC, and 40oC
|