The tradition of easy-to-use simulators continues in the Dale series. The DALE14 is powerful enough to test full function monitors with its onboard ECG, Arrhythmias, Respiration, Blood Pressure and Temperature simulations. The DALE14's pullout card is convenient for looking up all simulations for manual use.The DALE14 utilizes low profile ECG snap connectors that are AHA colour coded to connect to ECG snaps or clips. The user interface is simple to use with it's front panel softkeys. The unit comes with a User's manual, and adapter for lead testing and a battery eliminator
Features
- Compact
- 12-Lead ECG simulation
- Wide range of arrhythmias
- Auto/Manual operation
- RS-232 port
- Pullout card for quick menu access
Standard Accessories
- Operator's manual
- Battery eliminator
- Soft Vinyl Carrying Case
- Lead test adapter
Optional Accessories
- BP cables
- RS-232 cable
- Service manual
Specifications
ECG Waveforms
Normal Sinus rhythm: 30 - 350 BPM
Auto modePerformance Waveforms
Square wave: 2 Hz
Sine waves: 10, 40, 50, 60, 100 Hz
Triangle wave: 2 Hz
Pulse: 30, 60 BPM
Auto mode: 0.5 - 2.0 mVAplitudes
±2% lead II or ±5%
Arrhythmias
Arterial Fibrillation
Second degree heart block type 1
Right bundle branch block
Premature arterial contraction
Premature ventricular contraction early
Multifocul PVCs
Run of five (5) PVC's
Bigeminy
PVC R on T wave
Ventricular tachycardia
Ventricular fibrillation
Pacer
Fetal/matemal
Auto-sequence
Pulse wave: 4.0 s/1 mV
Sine wave bursts: 10, 40, 50, 60 and 100 Hz
Triangle wave: 2 HzHigh level output
0.5 V/mV, lead 1Respiration
Output configuration:
Lead I, II or RL to LL
Baseline impedance: 500 ohms or
1000 ohms, accuracy ±5 %
Delta Impedance: 0.1, 0.2, 0.5, 1.0,
and 3.0 ohms, accuracy ±10 %
Normal physiological rates: 0, 15, 20,
30, 40, 60, 80, 100 and 120 BPM
Apnea: 12 and 32 secondsBlood Pressure
Input/output impedance: 300 ohms
Output sensitivity: 5 or 40 µV/V/mmHg
Output range: 0 to +300 mmHgChannel 1 manually
selectable waveforms (mmHg)
Arterial = 120/80
Left ventricle = 120/0
Central venous pressure = 15/10
Right ventricle = 25/0
Pulmonary artery = 25/10
Pulmonary artery wedge = 10/2
Atmosphere (0)
Static: 20, 40, 80, 100, 200, 250, 300
Artifact BP1/BP2Automated Channel 1
blood pressure waveform
Sequence:
- Atmosphere (0)
- Static: 20, 40, 80, 100, 200 250 and 300Channel 2 manually
selectable waveforms (mmHg)
Atmosphere (0)
Track Bp1: duplicates channel 1 selection
Central venous pressure = 15/10
Swan-Ganz: Insert, Inflate, Deflate, RemoveAutomated Channel 2
blood pressure waveform
Sequence:
- Atmosphere (0)
- Static: 20, 40, 80, 100, 200, 250 and 300Temperature
30ºC (86ºF), 37ºC (98.6ºF), 40ºC (104ºF)Power
9 V alkaline batteryWeight
1 pound (0.4 kg)Dimensions
6.3 x 4.0 x 1.5 in