topbanner.GIF (2369 bytes)
bluedot.gif (128 bytes)products.gif (435 bytes) nav2.GIF (429 bytes) about.gif (349 bytes) contact.gif (392 bytes) news.gif (342 bytes)
 

Back to products list

Get a Quote

PMA-1 Pacemaker Analyzer


 

Test the complete performance of invasive and non-invasive external pacemakers, with one device.

OVERVIEW

The PMA-1 has been specifically designed for use with all types of external cardiac pacemakers including invasive and non-invasive (transthoracic) types. The comprehensive range of tests and simulations will meet the needs of all cardiac units and the print and recording features will achieve time and cost savings for the biomedical department.

The microprocessor-based analyzer enables measurement and display of pacing pulse rate, height and width as well as A-V interval, refractory period, R, S and T wave sensitivity, long term stability, and susceptibility to 50/60 Hz interference.

The internal programs generate the interference signals (50/60 Hz) as well as square, triangular, and haversine and ISO waveforms for the R, S and T wave simulations.

FEATURES

 

  • Includes all necessary test leads
  • Hand held portability
  • Large 2 line LCD
  • Comprehensive test ranges

     

SPECIFICATIONS
 

  • Display 2 line x 16 character LCD
     
  • Inputs
    Ventricle Pacing Signal: (0-12 v, 25 ma), (500 ohm load)
    Arterial Pacing Signal: (0-12 v, 25 ma), (500 ohm load)
    Transthoracic Pacing Signal: (0-40 v, 250 ma), (50 ohm load)
     
  • Outputs Parallel printer port
    Serial computer port
     
  • Measurement of Pacing Pulse
    Amplitude: 0-12 v invasive, 0-40 v non-invasive
    Rate: 30-300 bpm
    Pulse length: (<1-50 ms)
     
  • A-V Time Interval 0-1000 ms
     
  • 50/60 Hz Interference Signal 0-25 mv in 1 mv steps
     
  • Refractory Period Measurement 20-500 ms
     
  • Simulated R, S & T wave Width 5, 10..., 100, 200 ms
     
  • Waveforms Square, triangular, haversine, ISO Rate (30-120 bpm)
     
  • Amplitude 0-25 mv, in 0.5 mv steps
     
  • Long Term Tests Reference value for pulse rate, amplitude and length established from first 10 pulses. Thereafter any deviation over 10% (settable) flagged.
     
  • Power Requirements: 4 x AA cells; Current drawn 20 ma approx.; Or connect to 9 vdc supply
     
  • Dimensions: 7.5 x 5.5 x 1.5 in. (18.3 x 13.4 x 3.7 cm)
     
  • Weight: 1.54 lb (700 g approx.)
     
ACCESSORIES
 
  • Carrying case
  • RS232 cable
  • Printer cable
  • Parallel printer

    Back to products list