Modes of operation
» Line powered, battery
powered, offline
(battery maintenance
charge)
Test parameters
» Power (watts), HF/RF
current (mA), Test load
(ohm)
Selectable loads
» (monopolar and
bipolar)
» Range: 50-750 ohms
» Step size: 50 ohm
» Accuracy: +4% of
selection, 15 total
selections
Current measurement
(leakage)
» Range: 30-2500 mA
» Resolution: + 1 mA
» Accuracy: + 2.5% of
reading or + 15 mA
whichever is greater
Power measurement
(output)
» Range: 1 to 400 Watts
» Resolution: + 0.1
Watts
» Accuracy: + 5% of
reading or + 3 Watts,
whichever is greater
Frequency response
» Test load: 0 to 500
kHz
» System response: -3 dB
points, 1 kHz to 10 MHz
@ 300 Ohms
CQM test
» 50-750 ohms, 50 ohm
steps (see selectable
load for specs)
Oscilloscope output
» 10 to 1 transformer
coupled output,
uncalibrated
» Connector type: BNC
Battery
» Type: Sealed lead acid
» Time between recharge:
2 hours of continuous
use
» Time to full charge: 8
hours
» Number of cycles: 200
Top panel input
connectors
» Connector type: 4 mm
safety sockets
» Input voltage limit:
10,000 V peak
» Input current limit: 3
Amperes rms
Dimensions
» 6" H x 13" W x 10½" D
» (15.2 cm x 34.24 cm x
30.48 cm)
Weight
» 14.15 lbs. (5.6 kg)
|