   ESA 612


Overview | Features | Models | Standard Accessories | Optional Accessories | Specifications | Get A Quote
 




 

 
 
Overview

 

The ESA612 Electrical Safety Analyzer represents the next generation in test equipment for biomedical and clinical engineering professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.

With selection of two test loads, this versatile product can be used worldwide for compliance testing to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751) and AN/NZS 3551

The versatility of the multifaceted ESA612 tester is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.

 
Features


» Portable, ergonomic, lightweight, and easy to use
» Large, easy-to-read display with adjustable contrast
» Human-factors-designed user interface
» Tilt stand design for stand-up testing in field environments
» Five applied parts jacks and easy ECG snap connection with optional expander box
» ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool
» Replaceable mains fuses keep your unit in the field and out of the repair shop
» Internal memory for 100 test records
» 20 A at 120 V current capability
» USB connection for use with Ansur and Data Viewer software (for memory download to PC)
» Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center)
» Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
»
Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval

 

Models

 


» ESA612
 
Standard Accessories

» Operator's manual
 
Optional Accessories


» 
 
Specifications


Specifications
Voltage
Range (Mains Voltage) 90 V AC RMS to 132 VAC RMS
  180 V AC RMS to 264 VAC RMS
Range (Accessible Voltage) 0 to 300 VAC RMS
Accuracy ± (2 % of reading + 2 LSD)
Voltage Tests Mains and Point to Point
Earth Resistance
Two-Terminal Mode Test Current / Range & Accuracy >200 mA AC : 0.0 to 2.0 Ω ± (2 % of reading + 0.015 Ω)
Equipment Current
Mode AC RMS
Range / Accuracy 0 to 20 A : ± 5 % of reading ± (2 counts or .2A, whichever is greater)
Duty Cycle 15 A to 20 A, 5 min. on/ 5 min. off
  10 A to 15 A, 7 min. on/ 3 min. off
  0 A to 10 A continuous
Leakage Current
Modes * AC + DC (True RMS)
  AC only
  DC only
Patient load selection (Input Impedance) AAMI ES1-1993 Fig.1
  IEC 60601: Fig 15
Crest factor ≤ 3
Ranges 0.0 to 199.9 μA
  200 to 1999 μA
  2.00 to 10.00 mA
Frequency Response/ Accuracy DC to 1 kHz : ± (1 % of reading + 1μA or 1 LSD, whichever is greater)
  1 kHz to 100 kHz : ± (2 % of reading + 1μA or 1 LSD, whichever is greater)
  1 kHz to 5 MHz (current > 1.6mA): ± (4 % of reading + 1μA or 1 LSD, whichever is greater)
  100 kHz to 1 MHz : ± (5 % of reading + 1μA or 1 LSD, whichever is greater)
Leakage Tests Earth (ground wire)
  Chassis (enclosure)
  Patient (lead to ground)
  Patient Auxiliary (lead to lead)
  Mains on Applied Part (lead isolation)
  Direct Equipment
  Direct Applied Part
  Alternative Equipment
  Alternative Applied Part
  Accessible
  Point to Point
Mains on Applied Part 100 % of Mains
Differential Leakage
Ranges 75 to 199 µA
  200 to 1999 µA
  2.00 to 20.00 mA
Accuracy ± 10 % of reading ± (2 counts or 20 µA, whichever is greater)
Insulation Resistance
Ranges / Accuracy 0.5 MΩ to 20 MΩ : ± (2 % of reading + 2 counts)
  20 MΩ to 100 MΩ : ± (7.5 % of reading + 2 counts)
Source Test Voltage 500 V DC
  250 V DC
Insulation Resistance Tests Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part)
ECG Performance Waveforms
Accuracy ± 2 %
  ± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration
Waveforms / Rates ECG Complex (BPM) : 30, 60, 120, 180, and 240
Ventricular Fibrillation  
Square wave (50 % duty cycle) (Hz) 0.125 and 2
Sine wave (Hz) 10, 40, 50, 60, and 100
Triangle wave (Hz) 2
Pulse (63 ms pulse width) 30 and 60
Power Ratings
Mains Voltage Outlet 120 VAC : 230 VAC
Mains Voltage Inlet Power Range 90 to 132 VAC RMS : 180 to 264 VAC RMS
Maximum Current 20 A : 16 A
Hz 50 or 60 : 50 or 60
Physical Case
Dimensions (cm) 17.63 x 8.38 x 28.45
Dimensions (inches)  6.94 x 3.3 x 11.2
Weight 1.6kg (10.25 lb)
Environmental Specifications
Operating Temperature 10 to 40 degrees C
Storage Temperature -20 to 60 degrees C
Operating Humidity 10 to 90 % non-condensing
Altitude To 2,000 meters

* Modes are available in all leakage tests with the exception of MAP leakages that are available only in True RMS


 

Back to top

Back to products page