One-touch test selection. Three test loads and many
test standards to choose from.
The model 601 Pro SeriesXL is the most
advanced electrical safety analyzer for the
International market. The unique features of this
instrument will increase your productivity and
diversify your business to include laboratory
equipment. The one-touch testing user interface is
an industry exclusive that allows the user to
perform rapid tests on various medical devices
without having to manoeuvre around cumbersome menus.
The 601 Pro SeriesXL is the only
instrument that combines the EN 60601-1, EN
601010-1, and AAMI ES1 standard test loads into one
device.
Operational Description:
Voltage (Single and Dual Lead): Use this mode
of operation to troubleshoot electrical systems or
as a digital volt meter to verify the proper
installation of power systems. While in the IEC 1010
standard, a modified version of the dual-lead
measurement is performed with all measurements
referenced to earth and the red lead used as a
probe. The outlet powering the 601 Pro SeriesXL
is analyzed on start up and error messages are
displayed to notify the user what the status is of
the outlet power.
Insulation Resistance: Select between mains
or applied part insulation resistance testing. 500
VDC is applied for a minimum of 5 seconds in auto
mode and a maximum of 1 minute in step and manual
mode. Measured values are displayed in megOhms.
Protective Earth Continuity: Test current can
be selected by the user or selected automatically by
the assigned test standard. Choose between 1, 10 or
25 amps to comply with the Second Amendment of the
IEC-601. Measured values are displayed in milliOhms
along with the actual current applied.
Earth and Enclosure Leakage: No additional
connections are required when the Earth Leakage test
is selected, the 601 Pro SeriesXL
automatically inserts the measuring device into the
circuit. The user can then perform single fault
conditions. Enclosure leakage tests can be performed
with or without the applied parts grounded. Measured
values are displayed in RMS only.
Patient Leakage/Mains on Applied Part: Select
between 0, 1, 2, 3, or 5 (all) applied parts. All
applied part selections are made by the user or by
the device information record when called up.
Leakage measurements are displayed in RMS or DC
microamps. Mains on applied part generates a signal
that is at least 110% of incoming mains power and
applies it to the selected applied part. The signal
is applied with normal and reversed polarity.
Modes of Operation:
Auto: Utilize the standard specific
autosequences by selecting the auto mode. If the user
enters a control number that is resident in the
memory, the 601 Pro SeriesXL will
initiate a test based on the settings associated
with that stored control number. Failure to enter a
control number will result in the test progressing
but no data will be stored in memory. All tests are
compared to a preset table of limits and pass/fails
are applied.
Step: This mode will progress through an
auto-sequence but requires user interaction to move
from test to test.
Manual: This mode allows for the most
flexible testing sequence. The user can decide to
perform or skip any test available in a particular
standard. No data is stored but limits do apply in
this mode for printing purposes.