Overview
The
Impulse 6000D Defibrillator Analyzer and Impulse
7000DP Defibrillator/Transcutaneous Pacer

Analyzer
test systems
represent the
next generation in rugged, portable test
instrumentation for biomedical technical
professionals who need to ensure the proper
operation and ultimate performance of critical
life-support cardiac resuscitation equipment
The
Impulse 6000D and Impulse 7000DP test capabilities
encompass the spectrum of defibrillator pulse shapes
including Lown, Edmark, monophasic trapezoidal, DC
biphasic and pulsed AC biphasic. Both models feature
full-range AED test and simulation technology
compatibility and outperform in overall accuracy and
standards compliance.
These ergonomic test systems are rugged and easy to
carry, and both analyzers incorporate a first-rate
intuitive user interface with a backlit,
easy-to-read display and an internal rechargeable
high-capacity battery. The industry-standard 50-ohm
testload is built inside with optional test loads
available for in-depth device analysis. The Impulse
6000D and Impulse 7000DP feature unique ECG lead
attachment posts for safe and secure connections.
Integrated 12-lead ECG simulation capabilities with
independent outputs helps get the job done when
testing defibrillators, transcutaneous pacemakers
and ECG device performance including rate meter
accuracy and alarm verification.
Both models utilize DSP- (Digital Signal Process)
based technology to process waveforms and allow for
future firmware, software and
waveform upgrades. A standard USB interface allows
for computer control and data transfer; optional
Ansur PC-based automation software increases
productivity by outfitting users with an easy-to-use
method to standardize
testing procedures, capture waveforms and test
results, and print and document data to make the
Impulse 6000D and Impulse 7000DP the industry ideal
in defibrillator and transcutaneous pacemaker test
systems.
| |

|
» Monophasic,
biphasic, and pulsed-biphasic
energy measurement
» First-class measurement
accuracy
» Easy-to-use menu
» Internal dual-purpose (defib
and pacer) non-inductive (2 µH)
50 Ω testload
» Isolated outputs for better
measurements and low-noise ECG
simulation
» Both domestic and
international certifications:
CE, CSA, and C-tick (Australia)
» Protection of pacer input
against accidental application
of defibrillator output
» Internal high-capacity NiMH
battery
» Output energy, charge time and
synchronization defibrillator
tests
» Peak voltage and current
reading
» Full disclosure of monophasic
and biphasic defibrillator
waveform timing parameters
» Playback of defibrillator
output waveform
» 10 isolated ECG electrodes
provide 12 combinations for
standardized clinical signals
» Pacer current, voltage, pulse
rate, pulse width and energy
(Impulse 7000DP only)
» Test pacemaker pulse rate up
to 800 PPM (Impulse 7000DP only)
» Pacemaker sensitivity tests
(Impulse 7000DP only)
» Pacemaker refractory period
tests (Impulse 7000DP only)
» 60/50 Hz noise immunity test
(Impulse 7000DP only)
» Two-year factory warranty
» Optional Ansur PC-based
software plug-in available to
help standardize and document
testing procedures and results
» Designed, tested, and built to
incomparable Fluke quality
standards
|
|

» Impulse6000D
»
Impulse 700DP
|
|
|
|
| |

» Battery charger
» Operating/service manual
» Soft vinyl carrying case
|
|
|
» Hard-Sided
Carrying Case
|
|
|
|
| |

|
General Specifications |
|
General Specifications |
|
Temperature, operating |
10 °C to 40 °C (50 °F
to 104 °F) |
|
Temperature, storage |
-20 °C to +60 °C (-4
°F to +140 °F) |
|
Humidity |
10 % to 90 %
non-condensing |
|
Display |
LCD display |
|
Communications |
USB device port for
computer control |
|
Power |
Internal rechargeable
NiMH battery pack for
nine hours (typical)
operation after full
charge or the battery
charger can
simultaneously operate
the Analyzer and
charge the battery |
|
Mechanical |
Housing:ABS Plastic
Height:13 cm (5 in)
Width:32 cm (13in)
Depth:24 cm (9.5 in)
Weights (w/
batteries):3.02 kg
(6.6 lbs 0.1 oz) |
|
|
Defibrillator Analyzer
Specifications |
|
Compatible Defibrillator
Waveshapes |
| |
Lown, Edmark,
Trapezoidal, DC Bi-phasic,
and AC Pulsed Bi-phasic |
| |
AC Pulsed Biphasic
waveform has not been
approved in the United
States. |
|
|
Autoranged Measurement |
|
|
Accuracy |
|
0.1 J to 360 J |
±(1 % of reading + 0.1
J) |
|
360 J to 600 J |
±(1 % of reading + 0.1
J), typical |
| |
Note: For Pulsed Bi-Phasic
defibrillator, specified
accuracy is ±(1.5 % of
reading + 0.3 J) on both
ranges. |
|
|
Load resistance |
|
Resistance |
50 Ω |
|
Accuracy |
±1 %, non-inductive (<2
׀H) |
|
|
Pulse trigger level |
|
|
Pulse width |
|
Range |
1 ms to 50 ms |
|
Accuracy |
±0.1 ms |
|
|
Voltage |
|
Range |
20 V to 5000 V |
|
Accuracy |
±(1 % of reading + 2 V) |
|
|
Current |
|
Range |
0.4 A to 100 A |
|
Accuracy |
±(1 % of reading + 0.1
A) |
|
|
Sample rate |
|
|
Maximum Average Power |
| |
12 W, equivalent to 10
defib pulses of 360 J
every 5 minutes |
|
|
Scope Output |
|
Autorange |
2000:1, 400:1 and 80:1:
dependant on the range |
|
Waveform Playback |
Output: BNC |
| |
Output impedance: 50 Ω |
| |
Amplitude Accuracy:±5 % |
|
|
Charge Time Measurement |
|
Range |
0.1 s to 100 s |
|
Accuracy |
±0.05 s, typical |
|
|
Synchronization Test
(Elective Cardioversion) |
|
Delay Time Measurement |
Timing window: : ECG
R-wave peak to the defib
pulse peak |
| |
Range: -120 to +380 ms;
measures timing from 120
ms prior to the R-wave
peak to up to 380 ms
following the R-wave
peak |
| |
Resolution: 1 ms |
| |
Accuracy: ±1 ms |
|
ECG waves |
Normal Sinus Rhythm (NSR):
10 BPM to 180 BPM, in 1
BPM steps |
| |
Atrial fibrillation:
Coarse and fine |
| |
Monomorphic Ventricular
Tachycardia: 120 BPM to
240 BPM, in 5 BPM steps |
| |
Asystole: Flat line |
|
|
Automated Defibrillator
Test ECG Waves |
|
Normal Sinus |
10 BPM to 300 BPM, in 1
BPM steps |
|
Ventricular Fibrillation |
Coarse and fine |
|
Monomorphic Ventricular
Tachycardia |
120 BPM to 300 BPM, in 5
BPM steps |
|
Polymorphic Ventricular
Tachycardia |
5 types |
|
Asystole |
Flat line |
|
|
ECG Waves |
|
ECG
General |
|
Lead configuration |
12-lead simulation. RA,
LL, LA, RL, V1-6 with
independent signal
outputs |
|
Lead-to-lead impedance |
1000 Ω (nominal) |
|
Rate accuracy |
± 1 % of setting |
|
|
ECG
Amplitudes |
|
Reference lead |
Selectable, Lead II
(default) or Lead 1 |
|
Settings |
0.05 mV to 0.45 mV, in
0.05 mV steps hard
return 0.5 mV to 5.0 mV,
in 0.5 mV steps |
|
Accuracy (All
Performance waves and
Normal Sinus R waves) |
Lead II: ± 2 %
All other leads:
± 7 %
Defib paddles: ±
7 % |
|
Amplitude of ECG signals
relative to amplitude
setting (in percent) |
Lead II reference: |
| |
Performance waves and R
wave detection |
| |
Lead I – 70 % |
| |
Lead II – 100 % |
| |
Lead III – 30 % |
| |
Lead V1 – 100% |
| |
Lead V2 – 100% |
| |
Lead V3 – 100 % |
| |
Lead V4 – 100 % |
| |
Lead V5 – 100 % |
| |
Lead V6 – 100 % |
| |
Normal Sinus waves |
| |
Lead I – 70 % |
| |
Lead II – 100 % |
| |
Lead III – 30 % |
| |
Lead V1 – 24% |
| |
Lead V2 – 48% |
| |
Lead V3 – 100 % |
| |
Lead V4 – 120 % |
| |
Lead V5 – 112 % |
| |
Lead V6 – 80 % |
| |
Lead I reference: |
| |
Performance waves and R
wave detection |
| |
Lead I – 100 % |
| |
Lead II – 150 % |
| |
Lead III – 50 % |
| |
Lead V1 – 100% |
| |
Lead V2 – 100% |
| |
Lead V3 – 100 % |
| |
Lead V4 – 100 % |
| |
Lead V5 – 100 % |
| |
Lead V6 – 100 % |
| |
Normal Sinus waves |
| |
Lead I – 100 % |
| |
Lead II – 150 % |
| |
Lead III – 50 % |
| |
Lead V1 – 24% |
| |
Lead V2 – 48% |
| |
Lead V3 – 100 % |
| |
Lead V4 – 120 % |
| |
Lead V5 – 112 % |
| |
Lead V6 – 80 % |
|
|
ECG Normal
sinus |
|
Rates |
10 BPM to 360 BPM, in 1
BPM steps |
|
|
ECG High
level output (BNC Jack) |
|
Amplitude Range |
0.5 V/mV of reference
lead setting |
|
Accuracy |
± 5 % |
|
Output Impedance: |
50 Ω |
|
|
ECG on
defibrillator Input Load
|
|
Lead II reference |
40 % of amplitude
setting |
|
Lead I reference |
60 % of amplitude
setting |
|
|
ECG
Performance Waves |
|
Square wave |
2 Hz and 0.125 Hz |
|
Triangular wave |
2 Hz and 2.5 Hz |
|
Sine waves |
0.05, 0.5, 5, 10, 40,
50, 60, 100, 150, and
200 Hz |
|
Pulse |
30BPM and 60 BPM, 60 ms
pulse width |
|
|
R-Wave
detection |
|
Waveform |
Haver-triangle |
|
Rates |
30, 60, 80, 120, 200 and
250 BPM |
|
Widths |
8, 10, 12 and 20 to 200
(by 10) ms |
|
Amplitudes |
0.05 mV to 0.45 mV 0.05
mV hard return |
| |
0.5 mV to 5 mV, in 0.5
mV steps |
|
Accuracy |
± (1 % setting + 1 ms) |
|
|
Noise
Immunity |
|
Wave |
Sine |
|
Line Frequency |
50 Hz or 60 Hz ( ± 0.5
Hz) |
|
Amplitude |
Range: 0.0 mV to
10 mV, in 0.5 mV steps |
| |
Accuracy: ± 5 % |
|
|
Transvenous Pacer Pulse
Simulation |
|
Widths |
Range: 0.1, 0.2,
0.5, 1, and 2 ms. |
| |
Accuracy: ± 5 %
of setting |
|
Amplitudes |
Range: 0 (off)
and ± 2, ± 4, ±6, ±8,
±10, ±12. ±14, ±16, ±18,
±20, ±50, ±100, ±200,
±500, ±700 mV |
| |
Accuracy: ± (10 %
setting + 0.2 mV) |
|
Amplitude of Transvenous
Pacer Pulse Simulation
signals relative to
amplitude setting (in
percent) |
Lead II reference: |
| |
Lead I – 67 % |
| |
Lead II – 100 % |
| |
Lead III – 33 % |
| |
Lead V1 – 67% |
| |
Lead V2 – 67% |
| |
Lead V3 – 67 % |
| |
Lead V4 – 67 % |
| |
Lead V5 – 67 % |
| |
Lead V6 – 67 % |
| |
Lead I reference: |
| |
Lead I – 100 % |
| |
Lead II – 150 % |
| |
Lead III – 50 % |
| |
Lead V1 – 100% |
| |
Lead V2 – 100% |
| |
Lead V3 – 100 % |
| |
Lead V4 – 100 % |
| |
Lead V5 – 100 % |
| |
Lead V6 – 100 % |
|
|
Arrhythmia
selections |
|
Pacer Interactive (Transcutaneous
Pacer, Impulse 7000DP
only) |
Demand: 30 BPM to 360
BPM, in 1 BPM steps
Asynchronous
Non-Capture
Non-Function
Threshold (interactive
pacing simulation only):
10 mA to 250 mA in 10 mA
steps |
|
Supraventricular |
Atrial Fibrillation
Coarse
Atrial Fibrillation Fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
Paroxysmal Atrial
Tachycardia
Nodal Rhythm
Supraventricular
Tachycardia |
|
Premature |
Atrial PAC
Nodal PNC
PVC1 Left Ventricle (LV)
PVC1 Left Ventricle (LV)
Early
PVC1 Left Ventricle (LV)
R on T
PVC2 Right Ventricle
(RV)
PVC2 Right Ventricle
(RV) Early
PVC2 Right Ventricle
(RV) R on T
Multifocal PVCs (LV and
RV) |
|
Ventricular |
PVCs 6/min
PVCs 12/min
PVCs 24/min
Freq Multifocal
Trigeminy
Bigeminy
Pair PVCs
Run 5 PVCs
Run 11 PVCs
Monomorphic VTach: 120
to 300 (5 BPM steps)
Polymorphic VTach: 5
selections
Ventricular
Fibrillation: Coarse and
Fine
Asystole |
|
Conduction |
1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch
Block (RBBB)
Left Bundle Branch Block
(LBBB) |
|
Transvenous Pacemaker (ECG
simulation only) |
Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent
sinus beats
Demand with occasional
sinus beats
AV Sequential
Non-Capture
Non-Function |
|
|
Selectable
pacer pulse parameters for
transvenous simulation (Atrial
& Ventricular channels are
independently selectable by
the user) |
|
Atrial Pacer Pulse
(Note: relates only to
the TV pacer) |
Width: 0.1, 0.2, 0.5, 1,
2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to
20 (by 2), 50, 100, 200,
500, 700 mV |
|
Ventricular Pacer Pulse
(Note: relates only to
the TV pacer) |
Width: 0.1, 0.2, 0.5, 1,
2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to
20 (by 2), 50, 100, 200,
500, 700 mV |
|
|
(BNC Jack) |
|
Accuracy |
± 5 % for 2 Hz Square
Wave , 50 _ output
impedance |
|
|
Transcutaneous Pacemaker
Analyzer Specifications
(Impulse 7000DP only) |
|
Test Load
Selections |
|
Defibrillator Testload
(Internal) |
Fixed load: 50 Ω ±1 %,
non-inductive (<2 μH)
Power rating: 10 defib
pulses of 360 Joules
every 5 minutes |
|
Pacemaker Testloads
(Internal) |
Variable load: 50 to
1500 Ω in 50-Ωsteps (±1
%), non-inductive (<
2uH)
Power rating: 5 W
(average), 40 W (peak) @
1000 Ω |
|
|
Pacemaker
Measurements |
|
Manufacturer specific
algorithms |
GE Responder (1500 &
1700)
MDE 300 (Medical Data
Electronics)
Medtronic ERS/Physio
Control LIFEPAK
MRL (Medical Research
Laboratory/Welch Allyn)
Philips/Agilent HP
Schiller Medical
ZOLL Medical
(plus a general purpose
Default Algorithm
selection) |
|
Current |
4 to 250 mA, ± (1% of
reading + 0.02 mA) |
|
Pulse rate |
5 to 800 PPM, ± (0.5% of
reading + 0.1 PPM) |
|
Pulse width |
1 to 100 ms, ± (0.5% of
reading + 0.01 ms) |
|
Energy |
1 μJ to 2 J, ± (4% of
reading + 10 μJ) |
|
|
Demand and
Asynchronous Mode Test |
|
Input pacer pulse rates |
30 to 200 PPM |
|
ECG NSR wave |
Rate:10 to 300 BPM in 1
BPM increments.
Amplitude:1 mV
Underdrive rate:10 BPM
minimum
Overdrive rate:300 BPM
maximum |
|
|
Sensitivity Test |
|
Automatic interactive
threshold detection |
Compatible pacer rates:
30 to 120 PPM. |
|
ECG R Wave |
Waveforms:Square,
Triangle, Sine
Width:1 to 19 (by 1) ms.
20 to 95 (by 5) ms. 100
to 300 (by 25) ms
Amplitude:0.05 to 0.95
(by 0.05) mV. 1 to 5 (by
0.5) mV
Accuracy:± 5% of setting
|
|
|
Refractory
Period Tests |
|
Paced Refractory Period |
20 to 500 ms |
|
Sensed Refractory Period |
15 to 500 ms |
|
Accuracy |
± 1 ms |
|
Pacer pulse rate |
20 to 200 PPM |
|
ECG |
Waveform:Triangle wave
Pulse Width:40 ms
Amplitude:1 mV
|
|
Back to top
Back to products
list
|
|
|