Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator-Transcutaneous Pacer Analyzer

Overview | Features | Models | Standard Accessories | Optional Accessories | Specifications | Get A Quote

Overview

The Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator/Transcutaneous Pacer
An
alyzer test systems represent the next generation in rugged, portable test instrumentation for biomedical technical professionals who need to ensure the proper operation and ultimate performance of critical life-support cardiac resuscitation equipment

The Impulse 6000D and Impulse 7000DP test capabilities encompass the spectrum of defibrillator pulse shapes including Lown, Edmark, monophasic trapezoidal, DC biphasic and pulsed AC biphasic. Both models feature full-range AED test and simulation technology compatibility and outperform in overall accuracy and standards compliance.
 
These ergonomic test systems are rugged and easy to carry, and both analyzers incorporate a first-rate intuitive user interface with a backlit, easy-to-read display and an internal rechargeable high-capacity battery. The industry-standard 50-ohm testload is built inside with optional test loads available for in-depth device analysis. The Impulse 6000D and Impulse 7000DP feature unique ECG lead attachment posts for safe and secure connections. Integrated 12-lead ECG simulation capabilities with independent outputs helps get the job done when testing defibrillators, transcutaneous pacemakers and ECG device performance including rate meter accuracy and alarm verification.

Both models utilize DSP- (Digital Signal Process) based technology to process waveforms and allow for future firmware, software and
waveform upgrades. A standard USB interface allows for computer control and data transfer; optional Ansur PC-based automation software increases productivity by outfitting users with an easy-to-use method to standardize
testing procedures, capture waveforms and test results, and print and document data to make the Impulse 6000D and Impulse 7000DP the industry ideal in defibrillator and transcutaneous pacemaker test systems.

 
 
Features


» Monophasic, biphasic, and pulsed-biphasic energy measurement
» First-class measurement accuracy
» Easy-to-use menu
» Internal dual-purpose (defib and pacer) non-inductive (2 µH) 50 Ω testload
» Isolated outputs for better measurements and low-noise ECG simulation
» Both domestic and international certifications: CE, CSA, and C-tick (Australia)
» Protection of pacer input against accidental application of defibrillator output
» Internal high-capacity NiMH battery
» Output energy, charge time and synchronization defibrillator tests
» Peak voltage and current reading
» Full disclosure of monophasic and biphasic defibrillator waveform timing parameters
» Playback of defibrillator output waveform
» 10 isolated ECG electrodes provide 12 combinations for standardized clinical signals
» Pacer current, voltage, pulse rate, pulse width and energy (Impulse 7000DP only)
» Test pacemaker pulse rate up to 800 PPM (Impulse 7000DP only)
» Pacemaker sensitivity tests (Impulse 7000DP only)
» Pacemaker refractory period tests (Impulse 7000DP only)
» 60/50 Hz noise immunity test (Impulse 7000DP only)
» Two-year factory warranty
» Optional Ansur PC-based software plug-in available to help standardize and document testing procedures and results
» Designed, tested, and built to incomparable Fluke quality standards

 
Models

» Impulse6000D
» Impulse 700DP
 
 
Standard Accessories


» Battery charger
» Operating/service manual
» Soft vinyl carrying case

 

Optional Accessories


» Hard-Sided Carrying Case
 

 
 
 
Specifications


General Specifications
General Specifications
Temperature, operating 10 °C to 40 °C (50 °F to 104 °F)
Temperature, storage -20 °C to +60 °C (-4 °F to +140 °F)
Humidity 10 % to 90 % non-condensing
Display LCD display
Communications USB device port for computer control
Power Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can simultaneously operate the Analyzer and charge the battery
Mechanical Housing:ABS Plastic
Height:13 cm (5 in)
Width:32 cm (13in)
Depth:24 cm (9.5 in)
Weights (w/ batteries):3.02 kg (6.6 lbs 0.1 oz)

Defibrillator Analyzer Specifications
Compatible Defibrillator Waveshapes
  Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
  AC Pulsed Biphasic waveform has not been approved in the United States.
Autoranged Measurement
  0.1 J to 600 J
Accuracy
0.1 J to 360 J ±(1 % of reading + 0.1 J)
360 J to 600 J ±(1 % of reading + 0.1 J), typical
  Note: For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance 50 Ω
Accuracy ±1 %, non-inductive (<2 ׀H)
Pulse trigger level
  20 V
Pulse width
Range 1 ms to 50 ms
Accuracy ±0.1 ms
Voltage
Range 20 V to 5000 V
Accuracy ±(1 % of reading + 2 V)
Current
Range 0.4 A to 100 A
Accuracy ±(1 % of reading + 0.1 A)
Sample rate
  250 kHz (4 µs sample)
Maximum Average Power
  12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Scope Output
Autorange 2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback Output: BNC
  Output impedance: 50 Ω
  Amplitude Accuracy:±5 %
Charge Time Measurement
Range 0.1 s to 100 s
Accuracy ±0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement Timing window: : ECG R-wave peak to the defib pulse peak
  Range: -120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms following the R-wave peak
  Resolution: 1 ms
  Accuracy: ±1 ms
ECG waves Normal Sinus Rhythm (NSR): 10 BPM to 180 BPM, in 1 BPM steps
  Atrial fibrillation: Coarse and fine
  Monomorphic Ventricular Tachycardia: 120 BPM to 240 BPM, in 5 BPM steps
  Asystole: Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus 10 BPM to 300 BPM, in 1 BPM steps
Ventricular Fibrillation Coarse and fine
Monomorphic Ventricular Tachycardia 120 BPM to 300 BPM, in 5 BPM steps
Polymorphic Ventricular Tachycardia 5 types
Asystole Flat line

ECG Waves
ECG General
Lead configuration 12-lead simulation. RA, LL, LA, RL, V1-6 with independent signal outputs
Lead-to-lead impedance 1000 Ω (nominal)
Rate accuracy ± 1 % of setting
ECG Amplitudes
Reference lead Selectable, Lead II (default) or Lead 1
Settings 0.05 mV to 0.45 mV, in 0.05 mV steps hard return 0.5 mV to 5.0 mV, in 0.5 mV steps
Accuracy (All Performance waves and Normal Sinus R waves) Lead II: ± 2 %
All other leads: ± 7 %
Defib paddles: ± 7 %
Amplitude of ECG signals relative to amplitude setting (in percent) Lead II reference:
  Performance waves and R wave detection
  Lead I – 70 %
  Lead II – 100 %
  Lead III – 30 %
  Lead V1 – 100%
  Lead V2 – 100%
  Lead V3 – 100 %
  Lead V4 – 100 %
  Lead V5 – 100 %
  Lead V6 – 100 %
  Normal Sinus waves
  Lead I – 70 %
  Lead II – 100 %
  Lead III – 30 %
  Lead V1 – 24%
  Lead V2 – 48%
  Lead V3 – 100 %
  Lead V4 – 120 %
  Lead V5 – 112 %
  Lead V6 – 80 %
  Lead I reference:
  Performance waves and R wave detection
  Lead I – 100 %
  Lead II – 150 %
  Lead III – 50 %
  Lead V1 – 100%
  Lead V2 – 100%
  Lead V3 – 100 %
  Lead V4 – 100 %
  Lead V5 – 100 %
  Lead V6 – 100 %
  Normal Sinus waves
  Lead I – 100 %
  Lead II – 150 %
  Lead III – 50 %
  Lead V1 – 24%
  Lead V2 – 48%
  Lead V3 – 100 %
  Lead V4 – 120 %
  Lead V5 – 112 %
  Lead V6 – 80 %
ECG Normal sinus
Rates 10 BPM to 360 BPM, in 1 BPM steps
ECG High level output (BNC Jack)
Amplitude Range 0.5 V/mV of reference lead setting
Accuracy ± 5 %
Output Impedance: 50 Ω
ECG on defibrillator Input Load
Lead II reference 40 % of amplitude setting
Lead I reference 60 % of amplitude setting
ECG Performance Waves
Square wave 2 Hz and 0.125 Hz
Triangular wave 2 Hz and 2.5 Hz
Sine waves 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz
Pulse 30BPM and 60 BPM, 60 ms pulse width
R-Wave detection
Waveform Haver-triangle
Rates 30, 60, 80, 120, 200 and 250 BPM
Widths 8, 10, 12 and 20 to 200 (by 10) ms
Amplitudes 0.05 mV to 0.45 mV 0.05 mV hard return
  0.5 mV to 5 mV, in 0.5 mV steps
Accuracy ± (1 % setting + 1 ms)
Noise Immunity
Wave Sine
Line Frequency 50 Hz or 60 Hz ( ± 0.5 Hz)
Amplitude Range: 0.0 mV to 10 mV, in 0.5 mV steps
  Accuracy: ± 5 %
Transvenous Pacer Pulse Simulation
Widths Range: 0.1, 0.2, 0.5, 1, and 2 ms.
  Accuracy: ± 5 % of setting
Amplitudes Range: 0 (off) and ± 2, ± 4, ±6, ±8, ±10, ±12. ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, ±700 mV
  Accuracy: ± (10 % setting + 0.2 mV)
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference:
  Lead I – 67 %
  Lead II – 100 %
  Lead III – 33 %
  Lead V1 – 67%
  Lead V2 – 67%
  Lead V3 – 67 %
  Lead V4 – 67 %
  Lead V5 – 67 %
  Lead V6 – 67 %
  Lead I reference:
  Lead I – 100 %
  Lead II – 150 %
  Lead III – 50 %
  Lead V1 – 100%
  Lead V2 – 100%
  Lead V3 – 100 %
  Lead V4 – 100 %
  Lead V5 – 100 %
  Lead V6 – 100 %
Arrhythmia selections
Pacer Interactive (Transcutaneous Pacer, Impulse 7000DP only) Demand: 30 BPM to 360 BPM, in 1 BPM steps
Asynchronous
Non-Capture
Non-Function
Threshold (interactive pacing simulation only): 10 mA to 250 mA in 10 mA steps
Supraventricular Atrial Fibrillation Coarse
Atrial Fibrillation Fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
Paroxysmal Atrial Tachycardia
Nodal Rhythm
Supraventricular Tachycardia
Premature Atrial PAC
Nodal PNC
PVC1 Left Ventricle (LV)
PVC1 Left Ventricle (LV) Early
PVC1 Left Ventricle (LV) R on T
PVC2 Right Ventricle (RV)
PVC2 Right Ventricle (RV) Early
PVC2 Right Ventricle (RV) R on T
Multifocal PVCs (LV and RV)
Ventricular PVCs 6/min
PVCs 12/min
PVCs 24/min
Freq Multifocal
Trigeminy
Bigeminy
Pair PVCs
Run 5 PVCs
Run 11 PVCs
Monomorphic VTach: 120 to 300 (5 BPM steps)
Polymorphic VTach: 5 selections
Ventricular Fibrillation: Coarse and Fine
Asystole
Conduction 1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch Block (RBBB)
Left Bundle Branch Block (LBBB)
Transvenous Pacemaker (ECG simulation only) Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV Sequential
Non-Capture
Non-Function
Selectable pacer pulse parameters for transvenous simulation (Atrial & Ventricular channels are independently selectable by the user)
Atrial Pacer Pulse (Note: relates only to the TV pacer) Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
Ventricular Pacer Pulse (Note: relates only to the TV pacer) Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
(BNC Jack)
Accuracy ± 5 % for 2 Hz Square Wave , 50 _ output impedance

Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)
Test Load Selections
Defibrillator Testload (Internal) Fixed load: 50 Ω ±1 %, non-inductive (<2 μH)
Power rating: 10 defib pulses of 360 Joules every 5 minutes
Pacemaker Testloads (Internal) Variable load: 50 to 1500 Ω in 50-Ωsteps (±1 %), non-inductive (< 2uH)
Power rating: 5 W (average), 40 W (peak) @ 1000 Ω
Pacemaker Measurements
Manufacturer specific algorithms GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent HP
Schiller Medical
ZOLL Medical
(plus a general purpose Default Algorithm selection)
Current 4 to 250 mA, ± (1% of reading + 0.02 mA)
Pulse rate 5 to 800 PPM, ± (0.5% of reading + 0.1 PPM)
Pulse width 1 to 100 ms, ± (0.5% of reading + 0.01 ms)
Energy 1 μJ to 2 J, ± (4% of reading + 10 μJ)
Demand and Asynchronous Mode Test
Input pacer pulse rates 30 to 200 PPM
ECG NSR wave Rate:10 to 300 BPM in 1 BPM increments.
Amplitude:1 mV
Underdrive rate:10 BPM minimum
Overdrive rate:300 BPM maximum
Sensitivity Test
Automatic interactive threshold detection Compatible pacer rates: 30 to 120 PPM.
ECG R Wave Waveforms:Square, Triangle, Sine
Width:1 to 19 (by 1) ms. 20 to 95 (by 5) ms. 100 to 300 (by 25) ms
Amplitude:0.05 to 0.95 (by 0.05) mV. 1 to 5 (by 0.5) mV
Accuracy:± 5% of setting
Refractory Period Tests
Paced Refractory Period 20 to 500 ms
Sensed Refractory Period 15 to 500 ms
Accuracy ± 1 ms
Pacer pulse rate 20 to 200 PPM
ECG Waveform:Triangle wave
Pulse Width:40 ms
Amplitude:1 mV
 